Skip navigation View an alternate layout of this website with limited styles and no horizontal scrolling
Menu

The design and development of a gloveless endoskeletal prosthetic hand

By Doshi, Rajiv Yeh, Clement LeBlanc, Maurice; Journal of Rehabilitation Research and Development, Vol. 35, No. 4
(Pages: 8) Publication Date: October 1998

Current prosthetic hands, although functional, have the potential of being improved significantly. We report here the design and development of a novel prosthetic hand that is lighter in weight, less expensive, and more functional than current hands. The new prosthesis features an endoskeleton embedded in self-skinning foam that provides a realistic look and feel and obviates the need for a separate cosmetic glove. The voluntary-closing mechanism offers variable grip strength. Placement of joints at three locations (metacarpophalangeal and proximal and distal interphalangeal) within each of four fingers affords realistic finger movement. High-strength synthetic cable attached to the distal phalanx of each finger is used to effect flexion. A multiposition passive thumb provides both precision and power grips. The new prosthesis can securely grasp objects with various shapes and sizes. Compared to current hands, weight has been reduced by approximately 50%, and cable excursion required for full finger flexion by more than 50%. The new endoskeletal prosthesis requires approximately 12-24% less force input to grasp a variety of everyday objects, largely due to its adaptive grip. Production cost estimates reveal the new prosthesis to be significantly less expensive than current prosthetic hands.
Published by:
VA Rehabilitation Research & Development Service    (Web Site: http://www.rehab.research.va.gov )
Link to text: http://www.vard.org/jour/98/35/4/doshi354.htm

AbleData, 8630 Fenton Street, Suite 930, Silver Spring, MD 20910. 1-800-227-0216.
Maintained for the National Institute on Disability and Rehabilitation Research of the U.S. Dept. of Education
by ICF Macro under Contract No. ED-04-CO-0018/0007.

The records in AbleData are provided for information purposes only. Neither the U.S. Department of Education nor ICF Macro has examined, reviewed, or tested any product, device, or information contained in AbleData. The Department and ICF Macro make no endorsement, representation, or warranty express or implied as to any product, device, or information set forth in AbleData. The views expressed on this site do not necessarily represent the opinions of the Department of Education, the National Institute on Disability and Rehabilitation Research, or ICF Macro.